Ordering from outside of the United States?
Contact Paralan Corporation for information.
Get in touch
Description
The SPT-61 module is equipped with Pt 100 / Pt 500 / Pt 1000 type input with automatic recognition of 2- and 3-conductor connection. Due to precision of measurement 3-wire circuit is recommended. The measurement range spans from -100°C to +600°C, and input is fully linearised according to PN-EN60751+A2:1999 standard. Module allows conversion from temperature (RTD) to the current value in 4-20 mA standard. Measurement value is transmitted to the current loop output directly after recalculation according to scale selected by user. If input signal exceeds the permissible range, shortcut or break of measurement circuit occurs. The SPT-61 module is equipped with peaks detection function. It can detect peaks of the input signal and hold their values. The device and registers configuration may be done via build in USB interface, visible for PC as Virtual Com Port. Transducer parameters can be set by S-Config software. Every unit can be factory pre-configured on request.
Specifications
1
RTD
Pt100, Pt500, Pt1000: -100...+600°C
4...20 mA: passive current, isolated, operating range 3,4...24 mA max.
24V DC
mini USB
USB Virtual Communication Port
RS-232
9600 bit/s, 8N1 / 8N2
0...+50°C
IP 20
NORYL UL94 V0
DIN rail
6,1 x 80 x 93,8 mm
For more information on the SPT-61 please click the links below:
DATASHEET
MANUAL
**To order this product please select an option from each of the categories below
For more information, please visit this product's webpage.